Designed for continuous measurement of volume fraction of oxygen trace concentrations in various gases (nitrogen, argon, helium, hydrogen, propylene. etc.).
Application: cryogenic-liquid complexes, in chemical and metallurgy industries, in systems for process control.
Principle of operation - electrochemical.
Sampling – forced, by means of gauge pressure or exterior flow rate pump.
Version – general purpose industrial, for installation in non-explosive areas.
Variants of gas analyzer versions
Version |
Measuring/indicating range, ppm |
Least significant digit, ppm |
ИБЯЛ.413411.049 |
(0-5 000)/(0-50 000) |
1 |
ИБЯЛ.413411.049-01 |
(0-500)/(0-5 000) |
1 |
ИБЯЛ.413411.049-02 |
(0-100)/(0-1 000) |
0,5 |
ИБЯЛ.413411.049-03 |
(0-10)/(0-1 000) |
0,01 |
ИБЯЛ.413411.049-04 |
(0-10)/(0-1 000) |
0,01 |
ИБЯЛ.413411.049-05 |
(0-10)/(0-100) |
0,01 |
ИБЯЛ.413411.049-06 |
(0-2)/(0-100) |
0,01 |
ИБЯЛ.413411.049-07 |
(0-2)/(0-100) |
0,01 |
ИБЯЛ.413411.049-08 |
(0-2)/(0-100) |
0,01 |
Advantages:
- digital indication of measured value of oxygen volume fraction;
- light and sound alarms;
- change of threshold values via software;
- “dry” relay contacts for connection of and control over external devices;